Fluke 50
Series II
|
|
Thermometers
|
|
The
new Fluke 50 Series II contact thermometers offer fast response and
laboratory
|
|
accuracy (0.05% + 0.3๐C) in a rugged, handheld test tool.
|
|
Fluke
51 Seires II and 52 Series II product features:
|
|
Large
backlit dual display shows any combination of T1, T2 (52 only), T1-T2 (52
only),
plus MIN, MAX, or AVG
|
|
Relative
time clock on MIN, MAX, and AVG provides a time reference for major events
|
Electronic
Offset function allows compensation of thermocouple errors to maximize
overall accuracy
|
|
Measures J,
K, T, and E-types of thermocouples
|
|
Readout in ๐C, ๐F, or Kelvin
(K)
|
|
Splash and
dust resistant case protected by impact absorbing holster
|
|
User-friendly
front panel is easy to set up and operate
|
|
Sleep mode
increases battery life; typical 1000-hour battery life
|
|
Battery door allows easy battery
replacement without breaking the calibration seal
|
|
Optional ToolPak accessory allows the thermometer to hang from any
metal object
(with the rare earth magnet) or
secure around a pipe (with hook-and-loop straps)
for hands-free
operation.
|
|
3 year
warranty
|
|
Specifications
|
|
Temperature
|
|
Measurement
accuracy:
|
|
·
Above -100๐C (-148๐F)
|
|
J, K, T, E, and N-type**
|
+/-[0.05%
+ 0.3๐C (0.5๐F)
|
|
R** and S-type**
|
+/-[0.05%
+ 0.4๐C (0.7๐F)
|
|
·
Below -100๐C (-148๐F)
|
|
J, K, E, and N-types
|
+/-[0.20% + 0.3๐C (0.5๐F)
|
|
T-type
|
+/-[0.50% + 0.3๐C (0.5๐F)
|
|
Display
resolution:
|
0.1๐C/ ๐F/ K < 1000๐
|
|
1๐C/ ๐F/ K ≥ 1000๐
|
|
Measurement range:
|
|
J-type:
|
-210๐C to 1200๐C (-346๐F to 2192๐F)
|
|
K-type:
|
-200๐C to 1372๐C (-328๐F to 2501๐F)
|
|
T-type:
|
-250๐C to 400๐C (-418๐F to 752๐F)
|
|
E-type:
|
-150๐C to 1000๐C (-238๐F to 1832๐F)
|
|
N-type**:
|
-200๐C to 1300๐C (-328๐F to 2372๐F)
|
|
R** and
S-type**:
|
0๐C to 1767๐C (32๐F to 3212๐F)
|
|
Temperature
scale:
|
ITS-90
|
|
Applicable standards:
|
NIST-175
|
|
** Only the Fluke Models 53 and 54 Series II thermometers are
capable of measuring
|
|
N, R,
and S-type thermocouples.
|
|
Environmental
|
|
Operating
temperature:
|
-10๐C to 50๐C (14๐F to 122๐F)
|
|
Storage temperature:
|
-40๐C to 60๐C (-40๐F to 140๐F)
|
|
Humidity:
|
0%
to 90%; 0๐C to 35๐C (32๐F to 95๐F)
|
|
0%
to 70%; 0๐C to 50๐C (32๐F to 122๐F)
|
|
General
|
|
Unit weight:
|
400g (14 oz)
|
|
Dimensions:
|
17.3
cm L x 8.6 cm W x 3.8 cm D
(6.8
in L x 3.4 in W x 1.5 in D)
|
|
Battery:
|
3 AA batteries; typical 1000-hour life
|
|
Certification:
|
CE,
CSA, TÜV (pending)
|
|
Safety:
|
CSA C22.2 No. 1010.1 1992; EN 61010 Amendments 1,2
|
|
|
|
|
|
|