HANDY DIGITAL FORCE GAUGE HF SERIES
|
The HF
series can be used not only for simple measurement of tension/ compression
|
force
but also used for various testing purposes; such as switch test, connector
insertion/
|
withdraw test, and fracture test. Measured result and statistical
processed result can be
|
output for printing on your printer*
and also transmitted to your computer system.
|
*Text style printing by ASCII code.
|
Characteristics
|
ü
COMPARATOR FUNCTION
|
ü
MEMORY FUNCTION
|
Initial input of the tolerance limit
|
Measured results will be automatically memorized
|
(upper/lower) enables the device to judge
|
up to 256 measurements.
|
the measured result as GO/NG
|
ü STATISTICAL PROCESSING &
PRINTING OF
|
ü
VARIOUS OUTPUT
|
INSPECTION REPORT
|
Various options are offered for outputs;
|
The HF
series can calculate figures of the standard
|
CENTRONICS output, RS232C output,
|
deviation
and the process capability index and those
|
analog output, digimatic output and
|
figures
can be printed out on your printer in form of
|
switch output.
|
the inspection
report when the CENTRONICS interface
|
ü
MULTIPLE FUNCTIONS
|
(optional) is adopted.
|
Initial setting on the device is possible for
|
ü HIGH ACCURACY, HIGH
RESOLUTION
|
measurement of peak force, insertion/
|
The HF series have achieved extremely high accuracy;
|
withdraw force, insertion compression peak
|
such as ¡Ó0.2%FS of accuracy and
¡Ó0.1%FS of
|
force, withdraw extension peak force and
|
repeatable accuracy. And also,
the HF series perform
|
holding force at contact
point. Also, the device
|
extremely high resolution.
For example, the 10kgf
|
is applicable for setting of various printing
|
capable device can
indicate minimum 1gf. It means
|
modes, input of
local NEWTON
value and other
|
1/10000 resolution.
|
various modes settings.
|
ü LOLAL NEWTON VALUE INPUT
|
ü
AUTOMATIC POWER OFF
|
The
terrestrial gravitation varies with where
|
The HF series have automatic power-off function
|
measurement is performed. Measurement error by
|
to save battery and to keep the device
safe.
|
gravity
difference will be eliminated inputting the
|
power off time can be changed at the range of
|
local Newton value according
to the measurement
|
1-99 min.
|
place.
|
ü
EASY OPERATION
|
ü UNIT CONVERSION
|
Buttons are well arranged for easy operation to
|
It is possible to convert the graduations of N, kgf (gf)
|
perform multipurpose
measurement.
|
or lbf on
the device.
|
Specifications
|
HF-5
|
N
|
Kgf
|
gf
|
lbf
|
|
Max.Load
|
50.00
|
5.000
|
5000
|
11.02
|
|
Min. Resolution
|
0.01
|
0.001
|
1
|
0.01
|
|
ü
Measurement Unit:
|
N, Kgf, (gf),lb
|
|
ü
Accuracy:
|
+/-0.2%F.S.
|
|
ü
Repeatablility:
|
+/-0.1%F.S.
|
|
ü
Response Speed:
|
10~320HZ, 6 Steps
|
|
ü
Display:
|
4 Digit
LCD Display
|
|
ü
Measurement:
|
Maximum Value
Moment Value
|
|
ü
Sampling:
|
640
Times per sec.
|
|
ü
Allowable Load:
|
150% (Alarm Buzzer at 105%)
|
|
ü
A/D Converter:
|
A/D
converter 16bit
|
|
ü
Continuous Operation Time:
|
Approx. 10 hours (Full Charged at 25๐C)
|
|
ü
Charging Time:
|
Approx.
6 hours
|
|
ü
Operation Temperature:
|
0~+40๐C
|
|
ü
Power:
|
NICAD Battery, A/C Adaptor for Charging
|
|
ü
Weight:
|
500g
|
|
ü
Standard Accessories:
|
Attachment,
A/C Adapter, Cables, Case
|
|
|
|
|
|
|
|
|
|