Introduction
|
Portable
|
The
Impressor is a convenient tool for testing the
hardness of aluminum,
|
aluminum
alloys, copper, brass and other materials including plastics and
|
fiberglass. The instrument
is designed for use on fabricated parts and
|
assemblies as well as on raw
stock.
|
Easy to Use
|
No experience required; can be used in any position and in any
space that will
|
allow for
the operators hand. The hardness reading is instantly indicated on
|
the
dial, which is divided into one hundred graduations. No waiting,
|
pre-loading or
separate measurements.
|
Lightweight
|
The
impressor weighs only 1 lb. 2 oz. and comes
complete with carrying case,
|
adjusting wrench and two
spare indenter points, 2 lb. 8 oz.
|
Applications
|
Three
Models
|
GYZJ 934-1 for soft metals such as aluminum and its alloys,
brass, copper,
|
Available
|
And
some of the harder plastics and fiberglass. Approximate range 25 to 150
|
|
Brinell (10 mm ball 500
kg load). This unit meets American Society for
|
|
Testing
and Materials (ASTM) Standard D-2583.
|
|
GYZJ
935
for the softer plastics and very soft metals.
|
|
GYZJ
936
for extremely soft materials such as lead, linoleum and leather.
|
|
Corrosion
Control Products Company will be glad to recommend the most
|
|
suitable model upon
receipt of sample materials.
|
Operating Information
|
The
Impressor is best suited for testing homogeneous
materials. Materials of granular,
fibrous or
|
coarse
structure will produce a wide variation in hardness readings because of the
small diameter
|
of the indenter point.
|
For accurate readings, material should be at least 1-32 thick
and large enough for a minimum
|
distance of
1/8 in any direction from the indenter point to the edge of the specimen.
The testing
|
area
should be smooth and free from mechanical damage.
|
Simply
exert alight pressure against the instrument to drive the spring-loaded
indenter point into the
|
material. The indenter
point must be perpendicular to the surface being tested.
|
On very soft metals, the highest reading should be used since
cold flow permits the spring loaded
|
indenter
point to continue penetration.
|
Recommended Sample Sizes
|
Material
|
Hardness Scale
|
Reading Variance
|
Number of Readings
|
Variance of
Average
|
Homogeneous
Material
|
20
30
40
50
60
70
80
|
2.47
2.20
1.93
1.66
1.39
1.12
0.85
|
9
8
7
6
5
4
3
|
0.27
0.28
0.27
0.28
0.28
0.28
0.28
|
Reinforced
Plastics
|
30
40
50
60
70
|
22.4
17.2
12.0
7.8
3.6
|
29
22
16
10
5
|
0.77
0.78
0.75
0.78
0.75
|
|
|
|
|
|
|
|