MINIRAE®
|
3000+
|
Portable
Handheld VOC Monitor
|
|
Features and Benefits:
|
§ Most
advanced handheld VOC monitor on the market
|
§ PID
measures from 0-15,000 ppm
|
§ 3-second
response
|
§ Humidity
compensation
|
§ Patented PID
lamp auto-cleaning technology
|
§ Patented
Reflex PID Technology™
|
§ Optional
Mesh Radio for ConneXT compatibility
|
§ Optional
built-in Bluetooth transmitter that transmits real-time data up to 2 miles
with RAELink3 wireless router
|
§ Optional BLE
Module
|
§ Sample
pump draws from up to 100 feet
|
§ Correction
factors for more than 200 compounds
|
§ Built-in
flashlight
|
§ IP 65/67
water-proof and dust-tight
|
§ Logs up to
6 months of data at 1-minute intervals
|
§ Multi-language
capability – up to 10 languages
|
§ 9.8eV for
benzene and TVOC and 10.6 and 11.7 UV lamps for TVOC
|
|
Application:
|
§ Emergency
Response
|
§ Environmental
Cleanups/ Remediation
|
§ Hazmat
Response
|
§ Leak
Detection
|
§ Urban
Search & Rescue
|
§ Venue
Protection
|
§ Worker
Exposure
|
|
|
|
Specifications
|
Size
|
10”
L x 3.0”W x 2.5”H (25.5 cm x 7.6 cm x 6.4 cm)
|
Weight
|
26
oz (738 g)
|
Sensors
|
Photoionization
sensor with standard 10.6 eV or optional 9.8 eV or 11.7 eV lamp
|
Battery
|
§ Rechargeable, external
field-replaceable Lithium-Ion battery pack
§ Alkaline battery adapter
|
Running
time
|
16
hours of operation (12 hours with alkaline battery adapter)
|
Display
Graphic
|
4
lines, 28 x 43 mm, with LED backlight for enhanced display readability
|
Keypad
|
1
operation and 2 programming keys, 1 flashlight on/off
|
Direct
Readout
|
Instantaneous
reading
§ VOCs as
ppm by volume (mg/m³)
§ High
values
§ STEL and
TWA
§ Battery and
shutdown voltage
§ Date,
time, temperature
|
Alarms
|
95dB
at 12” (30 cm) buzzer and flashing red LEDD to indicate exceeded preset
limits
§ High: 3
beeps and flashes per second
§ Low: 2
beeps and flashes per second
§ STEL and
TWA: 1 beep and flash per second
§ Alarms
latching with manual override or automatic reseet
§ Additional
diagnostic alarm and display message for low battery and pump stall
|
EMC/RFI
|
Compliant
with EMC directive (2004/108/EC)
EMI and ESD test: 100MHz to 1GHz 30V/m, no alarm
Contact: +/-4kV
Air: +/-8kV, no alarm
|
IP
Rating
|
§ IP-67 unit
off and without flexible probe
§ IP-65 unit
running
|
Datalogging
|
Standard
6 months at one-minute intervals
|
Calibration
|
Two-point
or three-point calibration for zero and span.
Reflex
PID Technology™
Calibration
memory for 8 calibration gases, alarm limits, span values and calibration
dates
|
Sampling
Pump
|
§ Internal,
integrated flow rate at 500 cc/mn
§ Sample
from 100’ (30m) horizontally or vertically
|
Low
Flow Alarm
|
Auto
pump shutoff at low-flow condition
|
Communication
& Data Download
|
§ Download
data and upload instrument set-up from PC through charging cradle or using
BLE module and dedicated APP
§ Wireless
data transmission through built-in RF modem
|
Wireless
Network
|
Mesh
RAE Systems Dedicated Wireless Network
|
Wireless
Range (Typical)
|
Up
to 15ft (5m) for BLE
EchoView Host: LOS
> 660 ft (200 m)
ProRAE Guardian
& RAEMesh Reader: LOS > 660 ft (200 m)
ProRAE Guardian
& RAELink3 Mesh: LOS > 330 ft (100 m)
|
Safety
Certifications
|
US
and Canada:
CSA, Classified as Intrinsically Safe for use in
Class I, Division 1. Groups A, B, C, D
Europe: ATEX II
2G EEx ia IIC T4
|
Temperature
|
-4°
to 122°F (-20° to 50°C)
|
Humidity
|
0%
to 95% relative humidity (non-condensing)
|
Attachments
|
Durable
bright yellow rubber boot
|
Warranty
|
3
years for 10.6 eV lamp, 1 year for pump, battery, sensor and instrument
|
Wireless
Frequency
|
ISM
License-free band
IEEE
802.15.4 Sub 1 GHz
|
Wireless
Approvals
|
FCC
Part 15, CE R&TTE, Others
|
Radio
Module
|
Supports
BLE or Bluetooth or RM900
|
|
|
|