PRS-801 Resistance System Set
|
|
|
|
Resistance System Set
|
|
§ Automatic
Mode (Default): 0.01 volts to 10 volts Variable 1.0E-1 to 1.0E+4 Ohms
|
|
§ Accuracy:
Overall: +/-<5% at ambient conditions (at 23°C and 30% Rh)
|
|
§ Multi-function
2-5/8” x 1-5/8” Liquid Crystal Display with 0.5” digit height
|
|
§ Times
measurements in Seconds up to 99 seconds (Displayed on LCD)
|
|
§ Register
stores up to 80 data points (MEM # Displayed after RESET)
|
|
|
|
The
widest range portable constant voltage resistance instrument in today’s
market, the PRS-801
|
|
is the
choice of most ESD professionals. Why ? Because
the PRS-801 has an outstanding performance
|
|
combination of
measurement speed, wide range and accuracy.
|
|
|
|
Its
constant test voltage system is extremely stable and consistent with lab
level, bench top instruments.
|
|
|
|
Measurement Range
|
|
The
PRS-801’s wide range offers the maximum flexibility in measurement
applications. In effect, it is 3
|
|
instruments
in one because it function as:
|
|
1.
A low resistance range
instrument for measurements from 0.1 to 1.0x104 ohms. Most competitive
|
|
instruments do not measure
accurately at or below 1.0x103 ohms. Thus, the typical ESD practitioner
|
|
must carry a DMM for low resistance
checks of ground connections.
|
|
|
|
To ensure-field accuracy, the PRS-801CC Calibration Shunt is
used to confirm and adjust the accuracy
|
|
of its 0.1 to 10 ohm resistance
range.
|
|
|
|
2.
Wide range audit instrument to
meet ANSI/ESD S20.20 and TR53 procedures from 1.0x103 to 1.0x1011 ohms.
|
|
|
|
Measurement guidelines require accurate measurements from one
decade below to one decade above
|
|
the resistance requirements of a
facility’s ESD program. The PRS-801 easily exceeds this requirement by
|
|
as much as 3 decades, depending on
program requirements.
|
|
|
|
3.
Precision high resistance
measurement instrument for Packaging to S541 Standards, material and product
|
|
testing and acceptance verification.
Few competitive instruments can measure accurately up to or beyond
|
|
1.0x1011 ohms, while the PRS-801 comfortably measures up to
2.0x1014 ohms. There are no audit instruments
|
|
or laboratory bench top
instruments that perform in the PRS-801’s wide range.
|
|
|
|
Measurement Flexibility
|
|
Connect
to Virtually any 2-wire Fixture or electrode configuration.
|
|
|
|
The
PRS-801’s simple 3/4 inch terminal spacing provides connectivity to a variety
of electrodes and fixtures.
|
|
§ Terminal
design allows use of mechanically shielded or non-shielded banana plug leads
|
|
§ BNC
adapters with 3/4 inch spacing mount directly to the PRS-801 terminals for
measurements with wide range probes and fixtures
|
|
§ An
instrument reference allows use of a third wire instrument reference to
minimize effects of extraneous
fields
on precision fixture measurements
|
|
|
|
Measurement Speed: Electrification Period
|
|
The
PRS-801 is one of the fastest measurement instruments in today’s precision
measurement market. Laboratory
|
|
tests confirm
the PRS-801 can measure from 0 to 1.0x1012 ohms in 2.5 seconds. Thus, its
specified electrification
|
|
period (EP) is 8
seconds in accordance with ANSI/ESD STM 11.11 Surface Resistance of Planar
Materials.
|
|
Furthermore,
its EP is automatically adjusted to insure a stable measurement is displayed.
|
|
|
|
Specifications
|
|
Range
|
Resistance
from 0.1 (1.0E-1) Ohms to 200 Tera ohms (2.0E+14 ohms). Maximum Resistivity
with ANSI/ESD STM S11.11 concentric ring 2.0E+15 ohms/square.
|
|
Test
Voltage
|
Automatic
Mode (Default): 0.01 volts to 10 volts Variable 1.0E-1 to 1.0E+4 Ohms
|
|
Constant
Voltage
|
10
volts: +/-<0.2 volts 1.0E+4 to <1.0E+6 Ohms
100
volts: +/-<2.0 volts 1.0E+6 to 2.0E+14 Ohms
|
|
Manual
Mode
|
0.01
volts to 10 volts Variable 1.0E-1 to <1.0E+5 Ohms:10 Volts:+/-<0.2
volts 1.0E+2 to 1.0E+9 Ohms
100
volts:+/-<2.0 volts 2.0E+5 to 2.0E+14 Ohms
|
Accuracy
|
Overall:
+/-<5% at ambient conditions (at 23°C and 30% Rh).
|
|
Nominal
Range Tolerances
|
1.0E-1
to 1.0E+1 ohms: +/-5% Corrected for Test Lead Resistance
1.0E+1
to 1.0E+12 ohms: +/-2.0% with 10-foot Test Leads
1.0E+12
to 2.0E+14 ohms: <30% or +/-0.25 Decade with grounded, shielded leads
|
|
Display
|
Multi-function
2-5/8” X 1-5/8” Liquid Crystal Display with 0.5” digit height.
Display
3-1/2 digits in Ohms, or 1.0EXX in exponential format.
Ohms
Display Indicators: Ω, KΩ, MΩ, GΩ and TΩ. Includes 19-segment analog scale
(1-10 with 0.5 indication) with x1, x10, & x100
multipliers.
Number
of Data Points in Memory (0-80).
Automatic
Electrification Time (seconds), or Time required to
Manually obtain steady state measurement.
Displays
data HOLD, BATTERY status, MIN, MAX, AVG, REC and Test Voltage (<10, 10,
or 100 V)
|
|
LED
Indicators
|
14
Color LED’s from <103 to >1014 ohms.
Colors
(RED, GREEN, YELLOW/ORANGE, or Blank/OFF)
|
|
Timer
|
Times
measurements in Seconds up to 99 seconds (Displayed on LCD)
|
|
Memory
|
Register
stores up to 80 data points (MEM# Displayed after RESET)
|
|
RS-232
Output
|
Digital
format: Exponential Power followed by Integer
|
|