General Specifications:
|
|
Case
|
Polyethylene Copolymer, Chemical
Resistant, Light Weight &
Durable Temperature Range: -10๐F to
210๐F
|
Window
|
Lexan, Polypropylene, Impact Resistant
|
Chart Drive
|
C Battery Operated, UL and cUL Listed For Hazardous
Locations, Temperature Range: -40๐F to 160๐F
|
Chart Size
|
12-inch Diameter (American Meter Style)
|
Chart Rotation
|
60 Min, 92 Min, 12 HR, 1, 2, 7, 8, 14,
16, 31 & 32 Day
|
Number of Pens
|
1-3 (one for each element)
|
Pen Style
|
Disposable, Clip-On
|
Element Type
|
Pressure-Helical (bourdon), Stainless
Steel
Temperature-Helical (bourdon) Class V
Mercury Filled
|
Range
|
Pressure 0 thru 30,000 PSI, Temperature 0 250 Deg.F
|
Accuracy
|
+/- .5% of Full Scale
|
|
|
Operating Instructions
|
1) Thoroughly check instrument at the
time of unpacking to inspect for any damage that may have occurred during
shipment.
|
2) Mount the unit as level as possible.
|
3) Make sure the recorder is secure.
|
4) Select the desired rotation speed on the chart drive.
|
5) Install chart by unlocking the chart
hub.
|
6) Raise the Pen Lifter.
|
7) Slide the chart paper between the
pens and the chart plate securing it in the tabs.
|
8) Lock the Hub and lower the Pen Lifter.
|
9) Verify that the pen (s) are in in contact
with the chart paper and have sufficient ink.
|
10) Apply zero pressure and adjust the pen for chart zero.
|
11) Turn on the pressure to the unit.
|
12) Upon completion of test, turn off the Pressure.
|
13) Turn of the Chart Drive.
|
14) Lift pen (s) from Chart Paper.
|
15) Replace Pen covers to keep pens
from drying out.
|
|
|
|