This
hand-held instrument measures up to 100,000 rpm using infrared measurement.
|
A
combination digital/analog bargraph LCD displays
the measurement clearly and accurately.
|
Other
features include auto or manual range selection, smoothing, min/max recording
and
|
hold functions. An accessory kit of three mechanical
adapters allows for direct connection
|
and measurement to rotating shafts and other
devices.
|
The
Model CA25 includes a carrying case with 15 reflector strips, 9V Alkaline
battery and
|
a user manual.
|
Features
|
Non-contact
measurement through infrared beam not affected by ambient light
|
Contact
measurements with wheels (optional)
|
Rotation measurements to
100,000 rpm (and tr/min)
|
Frequency
measurements (Hz)
|
Linear speed
measurements (ft/min and m/min)
|
Periodic measurements
(ms)
|
% Duty cycle
|
Min/Max, Hold
and Smooth functions
|
Very high accuracy
(0.01%)
|
Applications
|
Motor rotation testing
|
Specifications
|
MODEL
|
CA25
|
RPM
|
Range
|
0 to 99,999 rpm*; 0 to 99,999 tr/min*
|
Accuracy
|
1.10-4
of Reading
+/-6cts
|
FREQUENCY (Hz)
|
Range
|
0.1 to 9999.9 Hz*
|
Accuracy
|
4.10-4
of Reading
+/-4cts
|
SPEED (Linear)
|
Range
|
0.6 to 59,999 m/min*; 2.0 to 99,999 ft/min*
|
Accuracy
|
m/min:
1.10-4 of Reading
+/-1ct;
ft/min:
3.10-4 of Reading
+/-1ct;
Sensor:
3.10-3
|
PERIODIC FUNCTION (ms)
|
Range
|
9999.9 to 0.1 ms*
|
Accuracy
|
1.10-4
of Reading
+/-5cts
|
% DUTY CYCLE
|
Range
|
0.2 to 125 Hz: 0.1 to
99.9%*; 125 to 500 Hz: 1 to 99%*
|
Accuracy
|
0.3 to 50
Hz: 0.1% of Scale; 50 to 125 Hz: 0.2% of Scale
125 to 500
Hz: 1% of Scale
|
EVENTS
|
Range
|
0 to 99,999 events @ 0.1
Hz to 10 kHz
|
Accuracy
|
+/-1 event
|
GENERAL
|
Other
Functions
|
Min/Max, Hold, Smooth
|
Sensors
|
Infrared
Optical Sensor (with reflective tape).
FRB input
connector* (for slow signals, +/-20V max,
1Hz to
1kHz and 0.1Hz to 10kHz on extended range)
|
Display
|
Extra large 95 x 55mm
LCD with bargraph
|
Dimensions
|
8.5 x 2.83
x 1.85 (216 x 72 x 47mm)
|
Weight
|
8.03 oz (250g)
|
Power Source
|
9V
Alkaline battery
|
Optional
Kit (Catalog#1749.02)
|
Calibrated wheel, cone
and cylinder for contact measurements.
Connects over optical
sensor.
|
*6 to 60rpm, 0.1 to 1Hz, 0.6 to 6m/min, 2 to 20ft/min, 1000
to 9999.9ms and 0.1 to 99%
(0.2 to 125Hz) through input connector only.
|