Product Features
::
|
Multi-Point Push-Button Calibration (Up to 3 points) for optimum
full-range accuracy of ± 2% full scale
|
::
|
Auto-Ranging for fast response and best resolution (up to
0.01 NTU) over wide measurement range
|
::
|
Push-Button Calibration is hassle-free, quick and easy
|
::
|
Auto-Alert Calibration prompts at user-defined interval to
ensure regularly scheduled recalibration for ISO and GLP compliance
|
::
|
RS232 Communication interface to download data to a printer or
PC for report documentation
|
::
|
Built-in Real-time Clock stamps stored data and calibration
data with time & date - meets standards for GLP purposes
|
::
|
Self-Diagnostic with message codes for easy troubleshooting
|
::
|
Pour-Through Assembly Module (optional) allows instantaneous sampling
without having to transfer samples into cuvettes
thus saving time.
|
|
The CyberScan TB1000 series turbidity meters (employing the Nephelometric non-ratio or NTU principle) are designed to meet
stringent requirements of today's scientists. TB1000 series meet the
performance criteria specified by USEPA method 180.1 for NTU measurement and
ISO 7027 (IR version). Ideal for beverage production,
drinking water, waste water treatment, petrochemical and electroplating
applications.
Specifications
Product Specification
|
Description
|
Turbidity Range
|
0 to 1000 NTU
|
Resolution & Accuracy
|
0.01 NTU (0 to 9.99 NTU); 0.1 NTU (10 to 99.9 NTU);
1 NTU (100 to 1000 NTU)
& ± 2% of Measured Range + 0.01 NTU
|
Repeatability
|
± 1% of Measured Range + 0.01 NTU
|
No. of Calibration Points
|
Up to 3 points
|
Calibration Options
|
4 points (0.02, 10, 100, 1000)
|
Light Source
|
White Light (Tungsten) or Infrared
|
Response Time
|
< 6 seconds
|
Real-time Clock
|
Date and Time
|
Output
|
RS232
|
Power
|
12VDC, using AC adapters 110VAC or 220VAC, 50/60 Hz
|
|