EJX430A
|
|
Gauge Pressure Transmitter
|
|
The high
performance gauge pressure transmitter EJX430A features single crystal
|
|
silicon resonant sensor and is suitable to measure liquid, gas,
or steam pressure.
|
|
The
EJX430A outputs a 4 to 20 mA DC signal corresponding
to the measured pressure.
|
|
It also
features quick response, remote setup and monitoring via BRAIN or HART
|
|
communications, and self-diagnostics. FOUNDATION Fieldbus
protocol type is also
|
|
available.
|
|
All EJX
series models in their standard configuration, with the exception of the Fieldbus
|
|
type, are certified by TUV as complying with SIL 2 for safety
requirement.
|
|
SPECIFICATIONS
|
|
SPAN AND RANGE LIMITS
|
|
Measurement
Span/Range
|
MPa
|
Psi(/D1)
|
Bar(/D3)
|
Kgf/cm2(/D4)
|
H
|
Span
|
2.5 to
500 kPa
|
10 to
2000 in H2O
|
0.025 to
5
|
0.025 to
5
|
Range
|
-100 to
500 kPa
|
-400 to
2000 in H2O
|
-1 to 5
|
-1 to 5
|
A
|
Span
|
0.0175
to 3.5
|
2.5 to
500
|
0.175 to
35
|
0.175 to
35
|
Range
|
-0.1 to
3.5
|
-14.5 to
500
|
-1 to 35
|
-1 to 35
|
B
|
Span
|
0.08 to
16
|
12 to
2300
|
0.8 to 160
|
0.8 to
160
|
Range
|
-0.1 to 16
|
-14.5 to
2300
|
-1 to 160
|
-1 to 160
|
|
|
PERFORMANCE SPECIFICATIONS
|
|
Zero-based
calibrated span, linear output, wetted parts material code S and silicone
oil,
|
|
unless otherwise mentioned.
|
|
For Fieldbus communication type, use calibrated range instead
of span in the following
|
|
Specifications.
|
|
NORMAL OPERATION CONDITION
|
|
(Optional
features or approval codes may affect limits.)
|
|
Ambient Temperature Limits
|
|
-40 to 85๐C (-40 to 185๐F)
|
|
-30 to 80๐C (-22 to 176๐F) with LCD display
|
|
Process
Temperature Limits
|
|
-40 to
120๐C (-40 to 248๐F)
|
|
Ambient Humidity Limits
|
|
0 to 100% RH
|
|
Maximum
Over Pressure
|
|
Capsule
|
Pressure
|
|
H, A
|
16 MPa (2300 psi)
|
|
B
|
25 MPa (3600 psi)
|
|
Working
Pressure Limits (Silicone oil)
|
|
Maximum
Pressure Limits
|
|
Capsule
|
Pressure
|
|
H
|
500 kPa (2000 inH2O)
|
|
A
|
3.5 MPa (500 psi)
|
|
B
|
16 MPa (2300 psi)
|
|
Supply & Load Requirements
|
|
With 24 V DC supply, up to a 550 W load can be used.
|
|
Communication
Requirements à
|
|
(Approval
codes may affect electrical requirements.)
|
|
BRAIN
|
|
Communication Distance
|
|
Up to 2 km (1.25 miles) when using CEV polyethylene-insulated
PVC-sheathed cables.
|
|
Communication distance varies depending on type of cable used.
|
|
Load
Capacitance
|
|
0.22 mF or
less
|
|
Load Inductance
|
|
3.3 mH or less
|
|
Input
Impedance of communicating device
|
|
10 kW or
more at 2.4 kHz.
|
|
|
|
|
|
|
|
|
|
|