ADM-870C AIRDATA MULTIMETER
|
|
AIR FLOW ● VELOCITY ● PRESSURE ●
TEMPERATURE
|
· Serial
output to printer or computer
|
· Laboratory
accuracy
|
· Mass flow
equivalent or true air flow and velocity
|
· 200
Reading memory with average/ total/ recall
|
· English
and metric
|
· Velgrid and
airfoil
|
· Supply and
exhaust
|
· Fume hoods
|
|
The ADM-870C AirData Multimeter is designed specifically for applications with
very demanding
|
air distribution requirements. The unique
features of this meter are ideally suited for maintaining air
|
quality control in facilities such as: hospitals,
pharmaceutical, biological and medical research
|
laboratories, semiconductor manufacturing clean rooms, and
other applications in which the
|
highest standards of the industry must
be met.
|
|
The
RS232 communications port allows the user to download readings to a
printer or computer.
|
Readings
can be downloaded either individually or as a sequence from memory. Optional
|
user-friendly software
allows readings to be automatically inserted into an Excel™ or similar
spreadsheet.
|
|
Air velocity and flow readings may be displayed as either standard
density mass flow equivalent,
|
or as local density air velocity or volumetric flow, as
compensated for variations in barometric
|
pressure and temperature.
|
|
The
auto-read function allows continuous automatic readings for monitoring
ongoing changes in
|
a system as
it is being adjusted.
|
|
Auto-read may also be used with the memory function to
automatically store up to 200 readings.
|
Each reading may be recalled, and the average and total of the
readings may be displayed.
|
The averaging feature is useful for measurements made under
fluctuating conditions.
|
|
The
differential pressure and temperature associated with a
velocity reading are in short-term memory
|
and may be
displayed immediately following each velocity measurement.
|
|
The multifunction model ADM-870C provides digital display of readings
in English and metric units
|
corrected for local or
standard density. Features include auto-read, memory with total
and average,
|
display of associated pressure and temperature, VelGrid, AirFoil probe and serial
output to printer or
|
computer. Measures air
velocity, absolute and differential pressure, temperature, and backpressure
|
compensated air flow when used
with the series 8400 FLowHood.
|
|
Specifications
|
AIR
VELOCITY:
+/-3% of reading +/-7 fpm from 50 to 8000 fpm using standard Series 160 pilot
tube
|
(29,000
fpm FS); 50 to 5000 fpm AirFoil; 50 to 2500 fpm VelGrid.
|
DIFFERENTIAL
PRESSURE:
+/-2% of reading +/-0.001 in wc from 0.0500 to
50.00 in wc, (0.0001 to 60 in wc
FS);
|
20
psid safe pressure.
|
TEMPERATURE: +/-0.5๐F accuracy
from 32๐F to 158๐F using
ADT440 Series TemProbes (-67๐F to 250๐F FS);
|
0.1๐F
resolution.
|
AIR
FLOW:
Accuracy is +/-3% of reading +/-7 cfm from 100 to 2000 cfm; range is 25 to
2500 supply,
|
25
to 1500 exhaust with 8400 FlowHood.
|
ABSOLUTE
PRESSURE:
+/-2% of reading +/-0.1 in Hg from 14 to 40 in Hg referenced to vacuum.
|
60
psia maximum safe pressure.
|
OPERATIONAL
TEMPERATURE LIMITS:
40๐F to 140๐F.
|
AIR
DENSITY CORRECTION:
Local or standard (mass flow) air density correction range is 14 to 40 in Hg
and
|
32๐F to 158๐F.
|
POSITION
SENSITIVITY:
Unaffected by position.
|
MEMORY: 200
readings, sequence labeled, sum and average.
|
CALIBRATION:
Calibration certified NIST traceable.
|
READ
OUT:
10 digit, 0.4, high contrast, liquid crystal display.
|
METER
HOUSING:
6.0 x 6.4 x 2.7 high impact ABS. 36 oz.
|
CONNECTIONS: 1/4"
OD slip-on for 3/16 ID soft tubing.
|
BATTERY
LIFE:
Up to 3000 readings/charge, 500 recharge cycles.
|