YSI Pro20i Dissolved Oxygen/Temperature
|
|
Portable Dissolved Oxygen Meter for the Field
|
|
|
|
Easy
to use, rugged and reliable, the YSI Pro20i provides everything you need in a
handheld
|
|
instrument for the
measurement of dissolved oxygen (DO). A true field instrument with IP-67
|
|
waterproof,
rubber over molded case, military spec connectors and 1-meter drop tests
ensure
|
|
durability to
provide years of sampling even in the harshest field conditions.
|
|
|
|
With
the Pro20i, accuracy is not compromised for ruggedness. The Pro20i can easily
be calibrated
|
|
for fast,
accurate measurements you can trust.
|
|
|
|
§ Integral
(i.e. built-in) cable with 1- 4- 10- m field cable
|
|
§ User-replaceable
DO sensors
|
|
§ One
Touch Cal allows for easy DO calibrations within 3 seconds with the press of
a button
|
|
§ Automatic
barometric pressure compensation
|
|
§ Quick
response times; T-95 response time in 8 seconds with standard membrane
(fastest response time on the market)
|
|
§ Stores 50
data sets; no need to write down data
|
|
§ Large
graphic, backlit display and glow in the dark keypad, visible in any lighting
condition
|
|
§ Display is
visible when wearing polarized sunglasses
|
|
§ A
true field instrument with IP-67, impact-resistant, waterproof case.
Waterproof with and without the battery cover.
|
|
§ Rubber
over molded, slip-proof case provides extra durability
|
|
§ Multiple
languages: English, Spanish, French, and German
|
|
§ Warranty:
3-year instrument, 2-year integral cable, 1-year polarographic
sensors, 6-month galvanic sensor
|
|
|
|
Pro20i
System Specifications (Instrument, Probe, and Cable)
|
|
Temperature
|
Range
|
23
to 131๐F (-5 to
55๐C) 0 to 45๐C for DO
compensation for mg/L
|
๐C, ๐F
|
Resolution
|
0.1๐F (0.1๐C)
|
|
Accuracy
|
+/-0.6๐F (+/-0.3๐C)
|
Dissolved Oxygen
|
Range
|
0
to 500%
|
%
Saturation
|
Resolution
|
0.1%
or 1% air saturation (user selectable)
|
%
Local
|
Accuracy
|
0
to 200% air saturation, +/-2% of the reading or +/-2% air saturation,
whichever is greater;
200
to 500% air saturation, +/-6% of the reading
|
Dissolved Oxygen
|
Range
|
0
to 50 mg/L
|
mg/L
|
Resolution
|
0.01
or 0.1 mg/L (user selectable)
|
ppm
|
Accuracy
|
0
to 20 mg/L, +/-2% of the reading or +/-0.2 mg/L, whichever is greater;
20
to 50 mg/L, +/-6% of the reading
|
Barometer
|
Range
|
400
to 999.9 mmHg
|
mmHg,
mbar, inHg,
|
Resolution
|
0.1
mmHg
|
psi,
kPa
|
Accuracy
|
+/-5
mmHg within +/-5๐C
temperature range from calibration point
|
|
|
|
|
|
|