Coating Thickness Measurement Instruments FMP10, FMP20, FMP30
and FMP40.
|
The
Flexible Solution for your Measurement Applications
|
|
Special
features
|
Fast and
non-destructive measurement on steel or iron (F) and non-ferrous metals (NF)
|
Automatic
probe and base material recognition
|
Large colour display
|
Supports
measurements according to IMO PSPC, SSPC-PA2, QUALANOD and QUALICOAT
|
USB interface,
Bluetooth or COM interface as option
|
Over
70 various high-precision probes for even the most sophisticated measurement
applications
|
|
DELTASCOPE®
FMP10 and FMP30
|
For
the measurement on ferrous base materials (F), e.g. paint, lacquer, powder
coating, chrome,
|
copper, zinc, as well as
enamel or plastic coatings on steel and iron.
|
|
Additional
features of the FMP30 and FMP40
|
Instrument features
|
External
key-triggered measurement acquisition, e.g. In hollow cylinders with small
diameters
|
Audible
and visual warning when tolerance limits are exceeded
|
Option Bluetooth
or COM additional available to the default USB interface
|
|
Measurement
application memory
|
Application memory
for up to 100 measuring applications incl. calibration (adjustment settings)
|
Memory for up to 20,000 readings
|
Allocation of
readings into up to 4,000 blocks
|
Date and time stamp for blocks
|
Application
linking mode: Common normalization/ calibration of measuring applications
|
Validation of the corrective
calibration by test measurements on standards
|
|
Measurement strategies and
evaluation
|
Stored
specifications for measurements according to IMP PSPC, SSPC-PA2, QUALANOD and
QUALICOAT
|
Capability
to enable matrix measurement mode for correlated multi-point measurements
|
Averaging of
measurement data: Only the mean value of several readings will be stored
|
Measurement
acquisition through area measurement: Single readings are captured until
probe lift-off and averaged
|
Outlier rejection
settings for automatic elimination of erroneous measurements
|
Free-running
display with additional presentation of the reading as an analog bar between
the tolerance limits
|
Statistics display
of the most significant values in the block and final results. Output of
variance-analytical values
|
Graphical
measurement display as a histogram
|
Capability of
entering process tolerance limits and computation of the associated process
capability indices Cp and Cpk
|
|
Probe selection based on several criteria
|
Material
combination of coating and base material
|
Thickness of coating and base material
|
Dimension of the
measurement area
|
Shape of the specimen
|
Surface condition
of the measurement area
|
|
Ordering Information
|
Order
no.
|
|
DELTASCOPE® FMP10
|
605-021
|
|
ISOSCOPE® FMP10
|
605-027
|
|
DUALSCOPE® FMP20
|
605-023
|
|
DELTASCOPE® FMP30
|
605-022
|
|
ISOSCOPE® FMP30
|
605-028
|
|
DUALSCOPE® FMP40
|
605-024
|
|