CA51/CA71 HANDY CAL calibrators are comprehensive generating/measuring
instruments
|
|
for all your calibration
and equipment checking needs at maintenance sites. The CA51/CA71
|
|
can
simultaneously generate and measure voltage, current, TC, RTD and pulse
signals, thus you
|
|
don’t need to prepare many
instruments. In addition, the incorporated rotary switches of good
|
|
repute
enable operations so smooth that an incorrect measurement caused from
erroneous
|
|
operations can be prevented.
|
|
|
|
General Specifications
|
|
Parameter
|
Specification
|
SOURCE
unit response time
|
Approximately 1 second (time between start of voltage change
and when voltage enters accuracy range)
|
SOURCE unit voltage
limiter
|
Approximately
32 V
|
SOURCE unit
current limiter
|
Approximately 25 mA
|
Divided output (n/m)
function
|
Output
= setting x (n//m) n=0 to m; m=1 to 19; n≤ m
|
Auto-step
output function
|
N value sent automatically when n/m function is selected
(two options: approximately 2.5 seconds/step or approximately
5 seconds/step)
|
Sweep function
|
Sweep
time (two options: approximately 16 seconds or approximately 32 seconds)
|
Memory
function
|
50 sets of data (generated and measured values are stored as
value sets with
the same address (up to 50 sets of data can be stored)
|
MEASURE unit maximum
input
|
Voltage
terminal: 300V DC/AC
Current
terminal: 120 mA DC
|
Current
terminal input protection
|
Fuses: 125 mA/250 V
|
MEASURE unit ground
voltage
|
Maximum
300 V AC
|
Measurement
display updating rate
|
Approximately once per second
|
Serial interface
|
Enables
when communication cable (RS-232) is connected;
sold
separately as optional accessory (CA71 only)
|
Display
|
Segment LCD (approximately 76 mm x 48 mm)
|
Backlight
|
LED
backlight; auto-off after one minute (from when LIGHT key is turned on)
|
Power
supply
|
Four AA-size (LR6) alkaline batteries, or special AC adapter
(sold separately)
|
Battery life
|
Measurement
off, output 5 V DC/10 kΩ or greater: Approximately 40 hours
Simultaneous
signal generation/measurement, output 5 V DC/10 kΩ or greater:
Approximately
20 hours
Simultaneous
signal generation/measurement, output 20 mA/5 V: Approximately 12 hours
(using
alkaline batteries, with backlight off)
|
Consumed
power
|
Approximately 7 VA (using 100 V AC adapter)
|
Auto-power-off function
|
Approximately
10 minutes (auto-power-off can be disabled through a DIP switch setting)
|
Applicable
standards
|
Standard
safety
|
IEC61010-1, EN61010-2-031
|
EMC
|
Emission
|
EN61236-1 Class B, EN61000-3-2, EB61000-3-3
EN55011: Group1, Class
|
Immunity
|
EN61326 Annex B
|
Insulation resistance
|
Across
input terminal and output terminal, 500V DC, 50 MΩ or greater
|
Withstand
voltage
|
Across input terminal and output terminal, 3.7 kV AC, for one
minute
|
Operating temperature
and humidity ranges
|
0
to 50๐C, 20 to 80% RH (no condensation)
|
Storage
temperature and humidity ranges
|
-20 to 50๐C, 90% RH or less (no
condensation)
|
External dimensions
(WHD)
|
Approximately
190 x 120 x 55 mm
|
Weight
|