TC 6621/ 6622
|
|
Handheld
temperature calibrator for thermocouples or resistive probes with memory
|
|
|
|
Easy-to-use
and equipped with a large graphical display, TC temperature calibrators are
the perfect
|
|
field tools for
easy and quick maintenance and commissioning of temperature transmitters and
probes.
|
|
|
|
TC 6621 measures or generates thermocouple temperatures (16
different types) and voltage at an
|
|
accuracy of 0.02%.
|
|
|
|
TC
6622 measures or generates RTD temperatures (12 different types) and resistance
at an accuracy
|
|
of 0.02%.
|
|
|
|
With
a very low temperature coefficient (10 ppm/๐C in
thermocouples and 7 ppm/๐C in resistance),
|
|
IP54
protection and robust design, TC calibrators are suitable for onsite use even
in demanding
|
|
environmental conditions.
They are widely used in the energy field, engineering sector, metal industry
|
|
and automotive
field.
|
|
|
|
TC
calibrators use a graphical interface making programming and reading easier,
under graphical or
|
|
test format.
Due to memory feature (10,000 values) and extended functionalities (square
root, steps,
|
|
synthesizer,
statistical functions
), TC are well adapted to different process job
procedures and ensure
|
|
full data
traceability as well as advanced data exploitation. Use them together with
DATACAL
|
|
calibration software
to display, manage data and issue your own customized calibration certificates.
|
|
|
|
TC
calibrators are delivered with 4 alkaline AA batteries in standard. The optional
battery charger used
|
|
a rechargeable
battery.
|
|
|
|
Key Features:
|
|
·
High accuracy: 0.02% reading with an
adjustable resolution of 1µV (TC
6621) and 1 mΩ (TC 6622)
|
|
·
Values displayed in ๐C, ๐F, mV and Ω
|
|
·
Low temperature coefficient: 10 ppm/๐C for
thermocouples and 7 ppm/๐C for resistance
|
|
·
Measurements with HOLD function
|
|
·
Simulation of ramps, preprogrammed steps and synthetizer values
|
|
·
Correction of calibrated sensors
|
|
·
Display of minimum, maximum and average value
|
|
·
Backlight
|
|
·
10,000 values stored and displayed graphically
|
|
|
|
Specifications
|
|
Specifications
and performances of TC 6621 @ 23๐C +/-5๐C
|
|
DC
voltage
|
|
Function
|
Range
|
Res.
|
Accuracy/1
year
|
Measuring
range
|
|
IN
|
100
mV
|
1
µV
|
0.020%
RDG + 3 µV
|
-10
mV/ 100 mV
|
|
OUT
|
80
mV
|
1
µV
|
0.020%
RDG + 3 µV
|
-9.5
mV/ 80 mV
|
|
Temperature
coefficient < 15 ppm/๐C beyond reference domain
|
|
|
|
General Specifications
|
|
Size
|
157
x 85 x 45 mm
|
Weight
|
306
g
|
Display
|
160
x 160 pixel liquid crystal graphical display with backlite
Display
of result as table of values or trend curve
|
Power
Supply
|
4
AA batteries 1.5V or rechargeable Ni-Mh batteries
with internal charger in option
|
Communication
Ports
|
USB
|
Storage
Capacity
|
10,000
data with date and time into one or several acquisition bursts
|
|
Environmental
Specifications
|
Reference
Range
|
23๐C +/-5๐C (RH: 45
to 75% w/o condensing)
|
Operating
Reference Range
|
-10
to 50๐C (RH: 20
to 80% w/o condensing)
|
Limit
Operating Range
|
-15๐C to +55๐C (RH: 10
to 80% w/o condensing) (70% at 55๐C)
|
Storage
Temperature Limits
|
-30๐C to +60๐C
|
Maximum
Height
|
0
to 2,200 m
|
IP
Protection
|
IP54
according to EN60529
|
|
|
|
|
|
|
|