General Specifications:
|
|
|
Case
|
Polyethylene
Copolymer, Chemical Resistant, Light
Weight & Durable Temperature Range:
-10๐F to 210๐F
|
|
Window
|
Lexan, Polypropylene, Impact Resistant
|
|
Chart
Drive
|
‘C’
Battery Operated, UL and cUL Listed For Hazardous
Locations, Temperature
Range: -40๐F to 160๐F
|
|
Chart
Size
|
12-inch
Diameter (American Meter Style)
|
|
Chart
Rotation
|
60
Min, 92 Min, 12 HR, 1, 2, 7, 8, 14, 16, 31 & 32 Day
|
|
Number
of Pens
|
1-3
(one for each element)
|
|
Pen
Style
|
Disposable,
Clip-On
|
|
Element
Type
|
Pressure-Helical
(bourdon), Stainless Steel
Temperature-Helical
(bourdon) Class V Mercury Filled
|
|
Range
|
Pressure
0 thru 30,000 PSI,
Temperature
0 – 250 Deg.F
|
|
Accuracy
|
+/-
.5% of Full Scale
|
|
|
|
|
Operating Instructions
|
|
1)
Thoroughly check instrument at the time of unpacking to inspect for any
damage that may have
occurred during
shipment.
|
|
2) Mount the unit as level as possible.
|
|
3)
Make sure the recorder is secure.
|
|
4) Select the desired rotation speed on the chart drive.
|
|
5)
Install chart by unlocking the chart hub.
|
|
6) Raise the Pen Lifter.
|
|
7)
Slide the chart paper between the pens and the chart plate securing it in the
tabs.
|
|
8) Lock the Hub and lower the Pen Lifter.
|
|
9)
Verify that the pen (s) are in in contact with the
chart paper and have sufficient ink.
|
|
10) Apply zero pressure and adjust the pen for chart zero.
|
|
11)
Turn on the pressure to the unit.
|
|
12) Upon completion of test, turn off the Pressure.
|
|
13)
Turn of the Chart Drive.
|
|
14) Lift pen (s) from Chart Paper.
|
|
15)
Replace Pen covers to keep pens from drying out.
|
|
|
|